Title (Primary) Sensor for amperometric determination of ammonia and ammonia forming enzyme reactions
Author Strehlitz, B.; Gründig, B.; Kopinke, H.
Source Titel Analytica Chimica Acta
Year 2000
Department UBZ; UBT_alt
Volume 403
Page From 11
Page To 23
Language englisch
Keywords amperometric enzyme sensor; polyaniline film; ammonia; screen printed electrode; urea
Abstract

The amperometric ammonia determination by use of a polyaniline-modified screen printed Pt–C electrode at 0.3 V (versus Ag/AgCl on the strip) is shown. The electrochemical pulse polarization method is used for the modification. The effects of basic electrode material, applied potential, pH of buffer solution and possible interferences of other ions were studied. The polyaniline-modified Pt–C electrode measures (NH4)2SO4 with a detection limit of 5 μM, a linear range up to 1 mM and a sensitivity of 40 ± 20 μA cm−2 mM−1. A new generation of enzyme sensors arises from the immobilization of ammonia-forming enzymes in a poly(carbamoyl)sulphonate hydrogel matrix on the working electrode as it is shown for urease to detect urea.

Persistent UFZ Identifier https://www.ufz.de/index.php?en=20939&ufzPublicationIdentifier=7693
Strehlitz, B., Gründig, B., Kopinke, H. (2000):
Sensor for amperometric determination of ammonia and ammonia forming enzyme reactions
Anal. Chim. Acta 403 , 11 - 23 10.1016/S0003-2670(99)00594-2