Details zur Publikation

Kategorie Textpublikation
Referenztyp Zeitschriften
DOI 10.1002/(SICI)1097-0231(19990715)13:13<1248::AID-RCM560>3.0.CO;2-H
Titel (primär) Standardization for oxygen isotope ratio measurement - still an unsolved problem
Autor Kornexl, B.E.; Werner, R.A.; Gehre, M.
Quelle Rapid Communications in Mass Spectrometry
Erscheinungsjahr 1999
Department ISOBIO; ANA
Band/Volume 13
Seite von 1248
Seite bis 1251
Sprache englisch
Abstract Numerous organic and inorganic laboratory standards were gathered from nine European and North American laboratories and were analyzed for their δ18O values with a new on-line high temperature pyrolysis system that was calibrated using Vienna standard mean ocean water (VSMOW) and standard light Antartic precipitation (SLAP) internationally distributed reference water samples. Especially for organic materials, discrepancies between reported and measured values were high, ranging up to 2‰. The reasons for these discrepancies are discussed and the need for an exact and reliable calibration of existing reference materials, as well as for the establishment of additional organic and inorganic reference materials is stressed.
dauerhafte UFZ-Verlinkung https://www.ufz.de/index.php?en=20939&ufzPublicationIdentifier=8131
Kornexl, B.E., Werner, R.A., Gehre, M. (1999):
Standardization for oxygen isotope ratio measurement - still an unsolved problem
Rapid Commun. Mass Spectrom. 13 , 1248 - 1251 10.1002/(SICI)1097-0231(19990715)13:13<1248::AID-RCM560>3.0.CO;2-H