Details zur Publikation

Kategorie Textpublikation
Referenztyp Zeitschriften
DOI 10.1016/S0003-2670(01)01030-3
Titel (primär) Rapid on-site determination of chlorobenzene in water samples using ion mobility spectrometry
Autor Borsdorf, H. ORCID logo ; Rämmler, A.; Schulze, D.; Boadu, K.O.; Feist, B.; Weiß, H. ORCID logo
Quelle Analytica Chimica Acta
Erscheinungsjahr 2001
Department ANA; GWS; MET; PB IBFL
Band/Volume 440
Heft 1
Seite von 63
Seite bis 70
Sprache englisch
Keywords Ion mobility spectrometry; Chlorinated hydrocarbons; Water analysis
Abstract

A fast screening procedure for water samples contaminated with chlorobenzene was developed. The volatile organic compounds were purged from water samples and their concentration was determined with ion mobility spectrometry (IMS) using corona discharge (CD) ionization. This analytical procedure permits the determination of chlorobenzene with a concentration range of 3–30 mg/l. The analytical results are available within 5 min. In order to test the general suitability of the method developed, the influence of sample matrix, especially of accompanying chlorinated aromatics, on ion mobility spectra of chlorobenzene was investigated.

dauerhafte UFZ-Verlinkung https://www.ufz.de/index.php?en=20939&ufzPublicationIdentifier=6220
Borsdorf, H., Rämmler, A., Schulze, D., Boadu, K.O., Feist, B., Weiß, H. (2001):
Rapid on-site determination of chlorobenzene in water samples using ion mobility spectrometry
Anal. Chim. Acta 440 (1), 63 - 70 10.1016/S0003-2670(01)01030-3