Titel (primär) A simple field method for determination of MTBE in water using hand-held ion mobility (IMS)
Autor Stach, J.; Arthen-Engeland, T.; Flachowsky, J.; Borsdorf, H. ORCID logo
Quelle International Journal for Ion Mobility Spectrometry
Erscheinungsjahr 2002
Department ANA; MET
Band/Volume 5
Heft 3
Seite von 82
Seite bis 86
Sprache englisch
Abstract The paper describes an analytical method for on-site detection of MTBE (Methyl tert-Butyl Ether, CAS 1634-04-4) using ion mobility spectrometry (IMS). The ion mobility spectra of MTBE show a typical dimer ion with a reduced ion mobility constant k0 = 1,50 cm2/Vs. Using this signal, it is possible to measure the contamination of MTBE in water in the range > 30 μg/l. Interference of matrix compounds like alkane’s, alkene’s and cycloalkane’s can be neglected due to their low proton affinity.
A simple and fast one-way sample preparation technique was developed for on-site analysis. For separation of MTBE from water and other volatile organic compounds (VOC’s), a water binding material (EXtrelut) in head space vials can be used. The gas phase above the EXtrelut is then injected in the hand-held IMS RAID-1 (Bruker Saxonia Analytik GmbH). The needed time for the determination of MTBE including sample preparation is about 5 minutes.
Mobile GC/MS (EM 640 from Bruker Daltonik GmbH) was used for verification of the IMS results. Detection limits below 1 μg/l MTBE were obtained with the same sample preparation procedures.