Details zur Publikation

Kategorie Textpublikation
Referenztyp Zeitschriften
DOI 10.1080/1025601031000147658
Titel (primär) Determining the content and 13C abundance of total dissolved carbon in water samples by TOC analyser-mass spectrometer coupling
Autor Russow, R.; Apelt, B.
Quelle Isotopes in Environmental and Health Studies
Erscheinungsjahr 2003
Department BOPHY
Band/Volume 39
Heft 4
Seite von 311
Seite bis 314
Sprache englisch
Abstract A combined system consisting of a TOC analyser connected to a quadrupole MS was recently described as a way of measuring the N content and the 15N abundance of total dissolved nitrogen in aqueous samples. This work examines whether this combination of instruments can also be used for the 13C determination of the total dissolved carbon in aqueous samples. A level of precision good for 13C-enriched samples was achieved with a relative standard deviation of <3%. By using an isotope ratio MS instead of the quadrupole MS employed here, TOC-MS coupling also ought to be suitable for determining natural 13C abundances.
dauerhafte UFZ-Verlinkung https://www.ufz.de/index.php?en=20939&ufzPublicationIdentifier=5172
Russow, R., Apelt, B. (2003):
Determining the content and 13C abundance of total dissolved carbon in water samples by TOC analyser-mass spectrometer coupling
Isot. Environ. Health Stud. 39 (4), 311 - 314 10.1080/1025601031000147658